Functions
|
Range
|
DC Voltage
|
0 to ±1000V (min.) : ±15ppm of Setting or better
|
DC Current
|
0 to ±20A (min.) (Continuous) : ±100 ppm of Setting or better
|
AC Voltage
|
0 -1000V : 10 Hz to100 KHz 1mV- 1000V(min.) : 10Hz to 500kHz (best)
±150 ppm of Setting or better
|
AC Current
|
30µA to 20A (min.) (continuous) : 10Hz to 30kHz(best), ±0.1% of Setting or better
|
Resistance
|
0 to 1GOhm (min.) : ± 30 ppm of Setting or better
|
Capacitance
|
220pF to 100mF (min.) : ±0.75% of Setting or better
|
Power (DC)
|
Voltage-33mV to 1000 V; Current- upto 20A; uncertainty- +/- 0.1%
|
Power (AC)
|
Voltage range- 33mV to 1000 V; Current Range- upto 20A
Uncertainty- +/- 0.1% of output at 20A
|
Frequency
|
0.01Hz to 1 MHz (min) Best Uncertainty : ± 25 ppm of setting or better
|
Thermocouple
|
Types: B, C, E, J, K, L, N, R, S, T & U and RTDs of all major types.
|
Analog Meter calibration
|
Compliance voltage from 4 V to 10V at different ranges
|
Display
|
6 ½ Digit LCD
|
Operating Power
|
230V ± 10%, 50/60Hz single phase; 47 to 63 Hz(min.)
|